ISO 9001:2015 certified environmental test systems
08Temperature simulation systems

Hot & Cold Test Chamber.

Controlled high- and low-temperature simulation for thermal cycling, endurance evaluation and product stress testing across components, materials and finished devices.

HeatingCoolingThermal cyclingRamp & soakProgrammable profiles
Capability range-70°C to +180°CPublished chamber introduction
Model upper limitUp to +200°CPublished ITHCC model table
Test-space volume125 L to 2500 LPublished model capacities
Resolution0.1°CTemperature display resolution
Chamber introduction

Thermal conditions made controlled and repeatable.

ISOTECH Hot & Cold Test Chambers simulate high- and low-temperature environments for evaluating product durability, reliability and performance. They support thermal cycling, endurance testing and stress simulation under accelerated or real-time conditions.

The chamber range is used for components, materials and complete devices serving automotive, defence, aerospace, electronics, renewable-energy and telecommunications programmes.

ISOTECH states that the systems are developed around requirements associated with IEC, IS, ASTM, MIL-STD and JSS testing frameworks. The chambers are manufactured in Bangalore and supplied in India and international markets with installation and customer support.

Specification contextThe page publishes a -70°C to +180°C general capability and model-table upper limits of +200°C. Confirm the selected model, working load and test profile with ISOTECH.
01 / Workspace125 L – 5000 LPublished main-feature range
02 / Temperature-70°C to +180°CPublished general range
03 / Accuracy±0.5°C to 2°CPublished temperature accuracy
04 / Resolution0.1°CTemperature resolution
05 / Gradient±1°C to 2°CPublished temperature gradient
06 / Rise & fall1°C to 15°CPublished average range

Where controlled temperature variation matters.

Thermal testing helps teams examine material stability, product endurance and performance across extreme operating conditions.

Automotive systems
Aerospace programmes
Electronics
Defence equipment
Renewable energy
Telecommunications
Construction / chamber bodyISOTECH Hot and Cold Chamber construction view
Chamber construction

Sealed, insulated and built for operating access.

The published chamber architecture combines a corrosion-resistant workspace with an insulated outer body, sealed access, viewing and specimen connections.

01

Interior workspace

Non-frost, vapour- and air-tight construction using 16 or 18 SWG, 304-grade stainless steel with adjustable and removable shelving.

02

External cabinet

Outer construction is available in 16 or 18 SWG CRCA, GI or stainless steel with electrostatic super-fine powder-coat finishing.

03

Thermal insulation

Asbestos-free and CFC-free mineral-fibre insulation supports heat retention and helps reduce operating loss.

04

Door and seal

A right-hinged front-opening single door uses double-lined grade-A silicone gaskets on both the door and chamber body.

05

Viewing and illumination

A multi-pane glass viewing window includes defogging heaters with automatic cut-off, together with illumination for observing the test specimen.

06

Ports and drainage

Left- or right-side cable ports with rubber plugs accept measurement and control connections. Drain outlet piping and suitable fittings are included.

Thermal performance

Three coordinated systems drive the temperature profile.

Air circulation, responsive heating and protected refrigeration work together to reduce spatial variation and maintain the programmed condition.

Air circulation

A suitably sized fan motor and impeller distribute conditioned air across the test space.

  • Impeller exposed inside the conditioning plenum
  • Drive motor remains outside the chamber
  • Detachable plenum cover supports maintenance
  • Published noise level below 70 dB

Heating system

Quick-response jacket-type air heaters bring the chamber toward the programmed hot condition.

  • Heater arrangement supports uniform temperature
  • No direct heater radiation onto the specimen
  • Reduced variation across multiple test items
  • Suitable for ramp-and-soak temperature profiles

Refrigeration system

Water- or air-cooled mechanical refrigeration is configured as single-stage or cascade compression according to the required range.

  • Hermetically sealed or boosted semi-sealed compressor options
  • Electronic automatic expansion valve
  • Water/air condenser and plate-fin cooler
  • R404A and R23 refrigerants listed in the current specification
  • Suction and head/discharge pressure measurement
  • HP/LP cut-off and oil-return separation
Instrumentation system

Programme the cycle. Review the thermal history.

Logic-programmable models use a large-screen controller with TFT LCD touch and key input, programme-pattern graphics, testing-history trends and cyclic ramp-and-soak segments.

TFT LCDRamp & soakTrend graphRS-232/485USB
01 / LO models

Programmable control

Multiple programmes and profile segments support ramp, soak and cyclic temperature operation.

02 / RE models

PID temperature control

RE models use a microprocessor PID non-programmable controller for temperature operation.

03 / Communication

Data and hardware links

High-speed RS-232/485, networking and USB communications support connected hardware and downloads.

04 / Measurement

PT100 sensing

A PT100 sensor measures chamber temperature. Power-failure recovery and password protection support controlled operation.

Safety devices

Operator, specimen and circuit protection.

Each functional electrical circuit has its own safety facility, capable of stopping the affected circuit or the entire chamber when a malfunction occurs.

Emergency stop switch
Control-circuit overcurrent protection
Thermal fuse and adjustable temperature cut-out
Door limit switch and stop-on-open protection
European-style signal tower
Independent circuit safety facilities
High- and low-pressure refrigeration cut-off
Chamber and specimen failure protection
ITHCC model matrix

Three minimum-temperature classes.

Published ITHCC logic-programmable models are differentiated by their low-temperature class and share an upper model-table limit of +200°C.

Electrical connection415 / 220 V AC ±10%, 3 / 1 phase / N / PE, 50 Hz
Model number Published temperature range Test-space volume
ITHCC-70 LO -70°C to +200°C 125 L to 2500 L
ITHCC-40 LO -40°C to +200°C 125 L to 2500 L
ITHCC-20 LO -20°C class to +200°C 125 L to 2500 L
Model key: LO denotes logic-programmable models. RE denotes non-programmable models.
Published-range note: The current page introduction and FAQ state -70°C to +180°C, while the model table lists upper limits of +200°C. Confirm the final operating range with the selected configuration.
Built chamber samples

Thermal systems in delivered form.

Published imagery from ISOTECH Hot & Cold Chamber projects.

Hot and Cold Chamber sample installation 1SAMPLE / 01
Hot and Cold Chamber sample installation 2SAMPLE / 02
Hot and Cold Chamber sample installation 3SAMPLE / 03

Published customer perspectives.

ISOTECH client reviews

Long-term use is described as robust, user-friendly and supported by a responsive technical team.

Radha KrishnaEnvironmental chamber customer

Product quality and customer service are highlighted positively in feedback on a salt-spray chamber project.

Sanjeev ShettySalt-spray chamber customer

Feedback identifies ISOTECH as an environmental-testing chamber manufacturer based in Bangalore.

LokanathPublished customer review

The supplied equipment is described in terms of reliability, durability and quality for environmental testing.

Malcom RoyPublished customer review
Define the thermal cycle

Share the high and low temperature profile.

Tell the ISOTECH team the specimen, range, workspace, ramp rate, dwell time, cycle count, electrical load and applicable test method.

Product questions